The FDA has issued an Early Alert after B. Braun told affected customers to remove certain hemodialysis bloodlines from where they are used or sold.

The affected device is the Streamline Bloodline Set for Dialog+, model SL-2010M2096, with UDI-DI codes 04046964367786 and 04046964367779. The impacted lots are A2600099, A2600128 and A2600136.

B. Braun said the required solvent application step during assembly was not performed, resulting in inadequate bonding of the dialyzer connector. As a result, the bloodline connector may detach or leak during priming or dialysis treatment, potentially leading to blood loss, clotting and occlusion of the extracorporeal circuit, and/or air ingress into the blood circuit.

As of October 10, 2026, B. Braun has reported no serious injuries or deaths associated with this issue.

Customers in the U.S. with questions about the recall should contact B. Braun at recalls@bbraunusa.com or (610) 849-1197.

"CDRH is issuing this 'Early Alert' to notify the public of a potentially high-risk device issue," the FDA said. "The FDA will keep the public informed and update this web page as significant new information becomes available."