The FDA has issued an Early Alert after Cardinal Health sent a letter to affected customers recommending updated use instructions for convenience kits containing infusion sets affected by the BD and CareFusion BD Alaris Pump Infusion Set Recall. Cardinal Health told customers that all convenience kits containing the affected infusion sets should be corrected before continued use.

According to the alert, BD said some Alaris Pump sets with restrictive components and designs — such as closed male luer, filters, check valves, Y-sites and microbore tubing — perform worse than previously disclosed. The deviations most commonly occur at low programmable rates under 1 mL/h and at or under 5 mL of "bolus volume."

BD has continued testing the infusion sets to update performance data for accuracy. That testing found the previously claimed performance was overestimated and, in some instances, the infusion performance is significantly worse. Performance attributes that differ from previously communicated values have the potential to result in greater degrees of under- or over-infusion, delayed time to occlusion alarms and bolus dose inaccuracies. The FDA said these performance issues may not be readily detected by users.

The FDA said the recalled device can lead to unpredictable pump behavior, such as under- or over-infusion or failure to infuse, which may lead to severe adverse events. The severity and nature of outcomes depend on various factors, but for the most vulnerable, high-risk populations — neonates, pediatrics and critically ill patients — the issues could result in serious adverse events or death, as these patients are more likely to receive lower rate and/or lower volume infusions, such as antiarrhythmics or blood-pressure supporting medications.

As of August 19, 2026, BD has not reported any serious injuries or deaths associated with this issue.

Customers in the U.S. with questions about the recall should contact Cardinal Health Quality Systems at (800) 292-9332, or their Cardinal Health sales representative.

The communication is part of the FDA's Medical Device Recalls and Early Alerts Program. "CDRH is issuing this Early Alert to notify the public of a potentially high-risk device issue," the FDA said. The agency said it will keep the public informed and update its web page as significant new information becomes available.